Activating the xBPI Protocol
Activate the xBPI protocol to perform adjustment with the service software
Sartocas software or Psion server without activating the BPI mode
through the SBI/BPI key.
Note: This menu item is accessible in the standard menu.
–Via Setup select the Device parameters menu item (for the COM1 port).
Open the Device parameters menu and scroll down to the Interface
submenu.
Open the Interface submenu and activate the xBPI menu item.
Note: Under xBPI a submenu is opened for assigning addresses, but this function is only
available when using xBPI with RS-485.
Display_xbpi.eps

Table of Contents

Other manuals for Sartorius combics 1

Question and Answer IconNeed help?

Do you have a question about the Sartorius combics 1 and is the answer not in the manual?

Sartorius combics 1 Specifications

General IconGeneral
BrandSartorius
Modelcombics 1
CategoryMonitor
LanguageEnglish

Summary

Overview

Service Concept

Prerequisites for performing maintenance and repair work on Sartorius Combics scales.

Overview of the Models

Auxiliary Service Tools and Equipment

Lists essential special tools required for servicing the Combics scale.

Operating Concept

Configuring the Combics

Function of the Service Keys and Switches

Activating the BPI Mode

Activating the xBPI Protocol

Activating the Service Mode

Working in the Service Mode

Menu Item 1-18: Entering Calibration/Linearization Weights

1-18-1: External User-defined Calibration Weight

Procedure for entering calibration weight without service mode.

1-18-2: Entering Linearization Weight

Procedure for entering linearization weights for the scale.

Calibration/Adjustment Routines in Service Mode (General Information)

Menu Item 1-9: Calibration/Adjustment Functions

1-9-3 External Calibration/Adjustment with User-defined Weights

Procedure for calibrating and adjusting using user-defined weights.

Repairing the Combics Indicator

Replacing the Power Cable

Replacing the Power PCB

Replacing the DC/DC Converter

Replacing the A/D Converter

Replacing the Digital PCB

Adjusting the Weighing Platform

Adjusting the Off-center Load

Procedure for adjusting the off-center load on four-load-cell platforms.

Setting the Overload Stops

Repairing the Weighing Platforms

Replacing Load Cells

Error Codes