29
– Press ), then, several times, k ) to select menu item 1-9-3, “External
calibration/adjustment with user-defined weights” and confirm by pressing )
briefly.
– Press ( to jump to the first columns of the numeric menu.
– Press ) to store the selected menu item.
– Turn the Combics indicator off and then on again to exit the Service mode.
– After taring or zeroing the scale, press and hold the ) key until Cal is
displayed.
If menu item 8-12-2 is active, the geographical data is displayed before the
calibration weight is prompted. Confirm each of the values shown by pressing
the ) key to continue with the calibration routine.
– After approx. 2 seconds the calibration weight (user-defined weight) required
is shown on the display. Select menu item “1-18-1” to enter the user-defined
weight.
– Place the prompted weight on the scale. After a brief pause, the difference
between the current value and the last span adjustment is displayed
(calibration).
– Press the ( key to stop the calibration/adjustment routine.
– Press the ) key to start the adjustment.
display_cal.eps
10Ib.eps
display_serv_01Ib.eps
10Ibw.eps
menu_193.eps