77
Diagram: Determining the
Adjustment Resistance
Note:
If a 125 lb test weight is used
(corresponding to the maximum
capacity of the weighing platform)
rather than a 12.5 klb test weight,
and the off-center load error is 0.8
lb, for example, then the same
adjustment resistance is set as
when a 12.5 klb test weight is used.
Diagram_usa.eps
Test load in klb
0,2123 45
2102030 4050
20 100 200 300 400 500
50lb 100lb 125lb 150lb 200lb 250lb 300lb
0,5klb 1klb 1,25klb 1,5klb 2klb 2,5klb 3klb
5klb 10klb 12,5klb 15klb 20klb 25klb 30klb
Corner load error in lb
Corner load balancing for 350 ohm load cells
(For 700-ohm load cells the computedvalue must be doubed)
Example: Test load: 12,5 klb
Corner load error: 80 lb
0,39+0,82+1,5+3
+1,5+3
1,5+3
0,39+0,82+3
0,82+3
0,39+3
3
0,39+0,82+1,5
0,82+1,5
0,39+1,5
1,5
0,39+0,82
0,82
0,39
Ω
.
0,39 Ω
Ω
Ω
Ω
Ω
Ω
Ω
Ω
Ω
Ω
Ω
Ω
Ω
.
.
.
.
.
.
.
.
.
.
.
.
Balancing resistence in ohm
Resistance
Individual
cholce