6 — Routine Maintenance
131
10432589_G AMSCO® 7000 Series
Operator Manual
Routine Maintenance
6.1 Preventive Maintenance
WARNING
ELECTRIC SHOCK AND/OR BURN HAZARD
● Fasteners and star washers are used to ensure protective bonding continuity. Always
reinstall any star washer which may have been removed during installation or
servicing.
● Pressing bottom portion of POWER (ON/OFF) toggle switch to turn off
washer/disinfector DOES NOT cut off electrical power. Lockout/Tagout building
electrical disconnect switch to turn off power to washer/disinfector for maintenance
or repairs.
● Repairs and adjustments to this equipment must be made only by STERIS or STERIS-
trained service personnel. Service personnel must disconnect all utilities to unit
before servicing. No one should service unit unless all utilities have been properly
locked out. Always follow local electrical codes and safety-related work practices.
WARNING
PERSONAL INJURY HAZARD
● Risk of pinch point between door and threshold when the door opens. Keep fingers
away from threshold.
● Risk of pinch point between door and upper panel. Do not push on top portion of
doors; do not push on door when door is rising; do not push on door when door is
jammed.