7 — Troubleshooting
200
Operator Manual
AMSCO® 7000 Series 10432589_G
7.83 ALARM# 34 CONDUCTIVITY TRANSMITTER DEFECTIVE (CRITICAL)
Condition
For EMEA only, when conductivity meter option is present: Alarm is triggered when signal from conductivity meter
appears defective. Conductivity meter or probe are not wired properly, conductivity meter is defective, or conductivity
probe is defective.
General Troubleshooting
Additional Information
Cycle must be aborted.
1. Press Silence Buzzer on touch screen to stop alarm
buzzer.
2. Press Ack on touch screen to acknowledge alarm.
3. Press Abort on touch screen to abort cycle.
4. Contact STERIS (Service charges may be incurred).
Note: if you need to continue running cycles while
waiting for service, the supervisor can disable
conductivity reading in Supervisor mode.
7.84 ALARM# 35 DRYING TEMPERATURE PROBES VARIANCE EXCEEDS RANGE (CRITICAL)
Condition
Alarm is triggered when drying temperature probe (RTD) readings are showing a variance over maximum allowed.
General Troubleshooting
Additional Information
Cycle must be aborted and then repeated.
1. Press Silence Buzzer on touch screen to stop alarm
buzzer.
2. Press Ack on touch screen to acknowledge alarm.
3. Press Abort on touch screen to abort cycle.
4. Repeat cycle.
5. If situation reoccurs, contact STERIS.
7.85 ALARM# 36 DRYING TEMPERATURE PROBE #2 DEFECTIVE (CRITICAL)
Condition
Alarm is triggered when during a cycle, signal from drying temperature probe #2 (TE2 [PV]) failed.
General Troubleshooting
Additional Information
Cycle must be aborted.
1. Press Silence Buzzer on touch screen to stop alarm
buzzer.
2. Press Ack on touch screen to acknowledge alarm.
3. Press Abort on touch screen to abort cycle.
4. Contact STERIS.