6 Routine Maintenance
136
Operator Manual
AMSCO® 7000 Series 10432589_G
6.4 Cleaning Water Outlet Connection
WARNING
BURN HAZARD
Inner surfaces of washer/disinfector are very hot after cycle is completed. Allow
washer/disinfector to cool down before touching chambers. Always wear appropriate
Personal Protective Equipment (PPE), including gloves as well as a face shield, and
avoid all contact with inner walls when reaching into chambers.
Once a week, allow washer/disinfector to cool down and open washer/disinfector door. Clean sliding water inlet as
follows:
1. Remove water sliding inlet and inspect for debris.
2. Brush off all surfaces (inside and outside) and rinse under tap water, as necessary.
3. Inner surface of water outlet connection within chamber sump must also be wiped.
4. Reinstall water sliding inlet.
5. Press Close Door on touch screen to close chamber load door.
1 Gasket
2 Water Outlet Connection
.
6.5 Cleaning Door Gasket
Once a month, allow washer/disinfector to cool down and open washer/disinfector door.
Then clean, brush and disinfect under both door gaskets using a concentrate germicidal detergent.