7 — Troubleshooting
167
10432589_G AMSCO® 7000 Series
Operator Manual
7.6 AUTO-DOCKING TRANSFER CART DOES NOT REMAIN DOCKED TO TRANSFER CART ADAPTOR
Condition
Auto-Docking Transfer Cart does not remain docked to Transfer Cart Adaptor.
General Troubleshooting
Additional Information
Auto-Docking Transfer Cart does not remain docked.
1. Refer to documentation provided with Auto-
Docking Transfer Cart.
7.7 CHAMBER DOES NOT DRAIN PROPERLY
Condition
Chamber does not drain properly.
General Troubleshooting Additional Information
Drain lines might be obstructed.
1. Visually verify if drain lines are obstructed.
2. If situation reoccurs, contact STERIS.
Drain extension time is too short
1. Set a longer drain extension time in Supervisor
mode.
2. If situation reoccurs, contact STERIS.
7.8 CHAMBER ROTARY SPRAY ARM ASSEMBLY DOES NOT ROTATE DURING CYCLE
Condition
Chamber rotary spray arm assembly does not rotate during cycle.
General Troubleshooting Additional Information
Rotary spray may be obstructed.
1. Verify if spray arm can move without any
constraint. If spray arm does not move freely, clean
spray arm hub (refer to Section 6.11, Inspecting and
Cleaning Rotary Spray Arm Assemblies AMSCO
7052HP W/D or Section 6.12, Inspecting and Cleaning
Rotary Spray Arm Assemblies - AMSCO 7053HP W/D
).
2. If situation reoccurs, contact STERIS.
There is debris in the spray arms.
1. Inspect if there is debris in spray arms, clean spray
arms if necessary (refer to Section 6.11, Inspecting
and Cleaning Rotary Spray Arm Assemblies AMSCO
7052HP W/D or Section 6.12, Inspecting and Cleaning
Rotary Spray Arm Assemblies - AMSCO 7053HP W/D
).
2. If situation reoccurs, contact STERIS.