7 Troubleshooting
174
Operator Manual
AMSCO® 7000 Series 10432589_G
7.26 LOAD COMES OUT DIRTY AT THE END OF CYCLE
Condition
Load comes out dirty at the end of a cycle.
General Troubleshooting Additional Information
Wrong type of detergent used.
1. Refer to detergent specifications and follow
detergent manufacturer's recommendations for
quantity of detergent to be used.
Rotary spray arms clogged.
1. Clean spray arm assemblies (refer to Section 6.11,
Inspecting and Cleaning Rotary Spray Arm Assemblies
AMSCO 7052HP W/D or Section 6.12, Inspecting and
Cleaning Rotary Spray Arm Assemblies - AMSCO
7053HP W/D ).
2. If condition reoccurs, contact STERIS.
Chamber debris screen clogged.
1. Clean debris screen (refer to Section 6.3, Cleaning
Debris Screen).
Sliding water inlet within chamber is not operating
properly.
1. Clean sliding water inlet and water outlet
connection.
2. If condition reoccurs, contact STERIS.