EasyManuals Logo

Steris AMSCO 7052HP User Manual

Steris AMSCO 7052HP
222 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #31 background imageLoading...
Page #31 background image
3 Component Identification
19
10432589_G AMSCO® 7000 Series
Operator Manual
1 Handheld Bar Code Reader
Figure 3-7. Handheld Bar Code Reader
3.3 Emergency Safety Features
Emergency Stop Pushbuttons
See and Figure 3-8.
The Emergency Stop pushbutton located under the control system on each side of the washer/disinfector instantly
stops washer/disinfector operation when pressed.
IMPORTANT: Use Emergency Stop pushbutton for emergency situations only. DO NOT USE as a Start/Stop button.
Figure 3-8. Emergency Stop Pushbutton
Chamber Door Interlock Mechanism
Chamber doors are equipped with a door interlock safety mechanism that prevents load side door from opening at the
same time as the unload side door to avoid cross-contamination. When a cycle is in progress, the door interlock
mechanism prevents either door from being opened without aborting the cycle.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Steris AMSCO 7052HP and is the answer not in the manual?

Steris AMSCO 7052HP Specifications

General IconGeneral
BrandSteris
ModelAMSCO 7052HP
CategoryWasher
LanguageEnglish