6 — Routine Maintenance
158
Operator Manual
AMSCO® 7000 Series 10432589_G
5. Press OK to confirm entry.
6. Display shows message requesting to verify both washer/disinfector doors are closed.
7. Press Start to begin test.
NOTE: Once test is completed, the 979-cycle reminder is reset to 0. Sump Safety Level Test reminder will be displayed
again in 979 cycles.
6.15 Water Inlet Pressure Test
IMPORTANT: Supervisor password is required to perform this procedure.
A water inlet pressure test is required to verify water inlet pressure and temperature in washer/disinfector.
1. Press Gear On Ready screen:
2. Press Supervisor mode on following screen: