7 — Troubleshooting
178
Operator Manual
AMSCO® 7000 Series 10432589_G
7.35 RETRIEVAL FAILED, NETWORK CONNECTION ERROR
Condition
Warning is displayed if CSiQ network is not responding.
General Troubleshooting
Additional Information
User needs to try again and continue without tracking if
fails.
1. Press RESTART to restart operator name retrieval
request.
2. Press CANCEL to return to cycle selection screen,
cancel cycle start and start over cycle selection.
3. Press CONTINUE to continue without tracking.
4. If the warning is displayed at every cycle, contact
STERIS.
7.36 RETRIEVAL FAILED, NO LOAD NUMBER READY
Condition
Warning is displayed if CSiQ did not respond any load number.
General Troubleshooting Additional Information
User needs to try again and continue without tracking if
attempt fails.
1. Press RESTART to restart load number retrieval
request.
2. Press CANCEL to return to cycle selection screen,
cancel cycle start and start over cycle selection.
3. Press CONTINUE to continue without tracking.
4. If the warning is displayed at every cycle, contact
STERIS.
7.37 RINSE RESERVOIR LEAKING
Condition
Warning is displayed if reservoir was refilled to many times in the last hour.
General Troubleshooting Additional Information
Water might be leaking from washer.
1. Press Silence Buzzer on touch screen to stop alarm
buzzer.
2. Press Ack on touch screen to acknowledge alarm.
3. Verify no water is on the floor.
4. If no water visible you can start a cycle as needed.
5. If water is visible, contact STERIS.