7 Troubleshooting
202
Operator Manual
AMSCO® 7000 Series 10432589_G
7.88 ALARM# 40 SOFTWARE APPLICATION ERROR
Condition
Error is triggered when and error occurred with washer software application.
General Troubleshooting
Additional Information
Error detected in application.
1. Press Silence Buzzer on touch screen to stop alarm
buzzer.
2. An error was detected in the application.
3. Press Ack on touch screen to continue, system will
reboot.
4. Washer/disinfector can be operated.
5. If error reoccurs, contact STERIS.
7.89 ALARM# 41 CHEMICAL CONTAINER [X] EMPTY
Condition
Alarm is triggered when no signal is detected from chemical [X] low-level sensor prior to initiating a cycle using that
chemical.
General Troubleshooting Additional Information
Pickup tube out of chemical container.
1. Press Silence Buzzer on touch screen to stop alarm
buzzer.
2. Press Ack on touch screen to acknowledge alarm.
3. Verify if chemical pickup tube is out of container. If
so, place pickup tube in chemical container.
4. If situation reoccurs, contact STERIS.
Chemical container empty.
1. Verify if chemical container is empty. If so:
a. Replace empty container with new one.
b. Prime chemical feed line to completely
remove air that entered the system.
c. Repeat cycle.
2. If situation reoccurs, contact STERIS.