7 Troubleshooting
179
10432589_G AMSCO® 7000 Series
Operator Manual
7.38 RINSE RESERVOIR NOT HEATING
Condition
Warning is displayed if reservoir temperature has increase of less than 9°F in the last 5 minutes.
General Troubleshooting
Additional Information
User can start a cycle as required.
1. Press Silence Buzzer on touch screen to stop alarm
buzzer.
2. Press Ack on touch screen to acknowledge alarm.
3. You can start a cycle as needed.
4. If the warning is displayed at every cycle, contact
STERIS.
7.39 SAVE COMPLETE SOME CORRUPT FILES WERE DISCARDED
Condition
If one or some files were corrupted, display shows message indicating that some files were deleted.
General Troubleshooting Additional Information
Files have been corrupted.
1. Press Ack on touch screen to acknowledge warning
and enter Cycle Selection menu.
2. Cycle graph files will be discarded.
3. If situation reoccurs, contact STERIS.
7.40 STATUS LIGHT DOES NOT TURN ON
Condition
Status light not turned on or defective.
General Troubleshooting Additional Information
Status light is burned out.
1. Contact STERIS.