7 — Troubleshooting
175
10432589_G AMSCO® 7000 Series
Operator Manual
7.27 LOAD COMES OUT WET AT THE END OF A CYCLE
Condition
Load comes out wet at the end of a cycle.
General Troubleshooting Additional Information
Custom cycle not programmed correctly.
1. Program Drying phase in custom cycle.
Drying type not properly selected.
1. Select proper Drying type in Supervisor mode.
Acu-Dry may not be appropriate for load processed.
1. Disable Acu-Dry feature.
Drying time is too short.
1. Extend treatment time.
2. If situation reoccurs, contact STERIS.
Acu-Dry may not be appropriate for load processed.
1. Disable Acu-Dry feature.
2. If situation reoccurs, contact STERIS.
Proper loading technique not used.
1. Reload rack.
HEPA filter clogged.
1. Contact STERIS.
Defective drying heating elements or drying heaters
circuit breaker tripped.
1. Look at temperature during Drying phase, by
pressing Details on touch screen, to confirm no
heating takes place. Contact STERIS.
7.28 PREVENTIVE MAINTENANCE DUE
Condition
Reminder is displayed when number of cycles set for next maintenance has been reached (Default is 10,000).
General Troubleshooting
Additional Information
User needs to acknowledge alarm and schedule
maintenance.
1. Press Ack on touch screen to acknowledge
reminder.
2. Contact STERIS to schedule a visit of a service
technician.
3. Reminder is displayed again after 20 more cycles
completed.
4. If reminder is displayed too often, ask service
technician to increase counter value.