1.5 INSTALLATION CHECKOUT
PROCEDURE
1.5.1 Purpose
This installation checkout procedure, when closely
adhered to, provides:
• Uniform procedures for all installations.
• Complete environmental data for future technical
assistance.
• Means to request Customer assistance in
environmental effects.
• Verification that SYSTEM 1E Processor is
functioning at maximum level.
• Data to track start-up reliability.
1.5.2 General Guidelines
An installation report is to be completed every time an
installation checkout is performed, a unit has been
moved to a new location or major change has been
made to the existing location.
NOTE: When installation checkouts are being
performed on more than one unit, a new installation
report must be completed for each SYSTEM 1E Liquid
Chemical Sterilant Processing System.
Refer to the applicable sections for definitions and use
of particular entries. If there is any confusion regarding
these instructions or completion of the installation
report, contact your supervisor for direction.
1.5.3 Distribution
There is only one copy to the installation report. The
installation report should be attached to the
corresponding field service report.
1.5.4 Procedure
The installation report form should be used to obtain
all information about the Customer and the
SYSTEM 1E Liquid Chemical Sterilant Processing
System at every installation checkout visit.
Complete the form in the following manner:
1.5.5 Electrical Utilities
1.5.6 Water Inlet
• CUST. NO: Enter complete Customer
number.
• CUST: Enter full name of Customer
facility
• MODEL NO: Enter model number of
SYSTEM 1E Liquid Chemical
Sterilant Processing System
(current model number is:
P6500 [115V ac, 60Hz])
• ADDRESS: Enter address of facility
where SYSTEM 1E Processor
is located.
• SERIAL NO: Enter serial number of
SYSTEM 1E Processor.
• GFI OUTLET,
GFI BREAKER
OR ISOLATED
POWER:
Enter the type of electrical
outlet: standard, GROUND
FAULT INTERRUPTED
outlet, GFI breaker or
ISOLATED POWER.
• VERIFIED
TRIP:
Check if GFI or (LIM) had
been observed to trip by
using a test button or GFI
tester.
•DUAL/
SINGLE
RECEPTACLE:
Circle one that applies.
IMPORTANT: Dual receptacle
required for UV light assembly.
• NEMA 5-20R/
OTHER:
Circle if NEMA series 5, 20
Amp, receptacle is provided
and is of hospital grade.
Describe any other
conditions.
• DEDICATED
LINE/OTHER:
Circle if the electrical outlet
is on a dedicated electrical
circuit and rated for 20 Amps
(115V ac). Describe any other
conditions.
• COMMENTS: Note any unusual conditions
or additional useful
information.
•DUAL
PREFILTER:
Dual prefilter configuration
is standard. No optional
configurations available.