4-2
764333-675
Flexible Tray for Processing: Wipe inner surface
and seal, tray for processing, and accessory rack
with a soft cloth dampened with 70% isopropyl
alcohol (Figure 4-4).
4. Wipe aspirator assembly and liquid chemical
sterilant compartment with a soft cloth dampened
with 70% isopropyl alcohol (Figure 4-5).
5. Check aspirator assembly. Verify probe lumen is
clear. Inspect aspirator for cracks or chips. Verify
hose connections are secure.
Important: If any damage is visible on the aspirator
assembly, replace it immediately.
WARNING LIQUID CHEMICAL
STERILANT VERIFICATION HAZARD:
Use of a blocked or damaged aspirator
may result in ineffective liquid chemical
sterilization.
6. Check Drain Screen (Figure 4-6)
a. Drain screen is located in the bottom of sterilant
compartment.
b. Verify screen is clean. Remove any debris or lint
from screen. (A straight hemostat or smooth
pickups helps.)
c. Screen may be removed for thorough cleaning
by lifting processing tray and unscrewing
drogue. Replace screen after cleaning.
7. Remove processing tray from drip pan.
8. Clean drip pan (see Figure 4-7).
Clean drip pan using a soft cloth dampened with
70% isopropyl alcohol.
Figure 4-4. Clean Flexible Tray
Figure 4-5. Clean Flexible Tray
!
Figure 4-6. Check Drain Screen
Figure 4-7. Clean Drip Pan