4-1
764333-675
4.1 General
Maintenance procedures described in this section
should be performed at regular intervals, as indicated.
The frequency indicated is the minimum, and should
be increased if usage of the unit demands. Should a
problem occur, refer to S
ECTION 5, TROUBLESHOOTING.
4.2 Daily Maintenance of SYSTEM 1E
™
Liquid Chemical Sterilant Processing
System
WARNING – PERSONAL INJURY AND/OR
EQUIPMENT DAMAGE HAZARD:
Disconnect all utilities to SYSTEM 1E
Liquid Chemical Sterilant Processing System
before servicing. Do not service the SYSTEM 1E
Liquid Chemical Sterilant Processing System
unless all utilities have been properly locked out.
Always follow OSHA Lockout-Tagout and
electrical safety-related work practice standards.
1. Clean external surfaces (Figure 4-1).
Wipe SYSTEM 1E Processor with a soft cloth
dampened with 70% isopropyl alcohol.
2. Open lid (Figure 4-2).
3. Clean Tray for Processing (Figure 4-3).
Tray/container for Processing: Wipe inner
surface and seal, tray/container for processing, and
accessory rack with a soft cloth dampened with 70%
isopropyl alcohol.
Figure 4-1. Clean SYSTEM 1E Processor
External Surfaces
Figure 4-2. Open Lid
Figure 4-3. Clean Tray for Processing
Section 4: Inspection and Maintenance