EasyManuals Logo
Home>Steris>Laboratory Equipment>SYSTEM 1E

Steris SYSTEM 1E Service Manual

Steris SYSTEM 1E
179 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #57 background imageLoading...
Page #57 background image
5-5
764333-675
5.1.7 Concentration Low
Description – Control measures less than 175 counts
during exposure.
Action:
1. No S40™ Sterilant Concentrate – run cycle again
with S40 Sterilant Concentrate.
2. Sterilant builders/buffers hardened – run cycle
again with fresh S40 Sterilant Concentrate.
3. Scale build-up on CP1 and CP2 – clean.
4. Faulty CP1 and CP2 connection – repair.
5. Out of calibration – recalibrate.
6. Faulty control board – replace.
7. Faulty interface board – replace.
5.1.8 Chamber Level Low
Description – SYSTEM 1E Processor senses a not-full
condition during exposure.
Action:
1. Leaking V5 – replace pinch valve sleeve.
2. Faulty SOL4 – repair. (Refer to Figure 5-5.)
3. Faulty LS3 – replace. (Refer to Figure 5-3.)
4. Leaking tray – replace.
5. CK2 or CK3 leaking replace.
6. Faulty CK8 – rebuild or replace.
7. Leaking lid – replace.
8. Plumbing leak – repair.
5.1.9 Heater Problem
Description – Control senses temperature difference
of more than 7°C between TC1 and TC2.
Action:
1. Clean tray drain strainer.
2. Faulty P1 – replace. (Refer to Figure 5-5.)
3. Faulty SSR4 – replace. (Refer to Figure 5-5.)
4. Faulty TC1 or TC2 – repair or replace. (Refer to
Figure 5-2.)
5. TC1 or TC2 out of calibration – recalibrate.
6. Faulty interface board – replace.
7. Faulty control board – replace.
8. Processor may have been too cold if first cycle of the
day run cycle again.
5.1.10 Chamber Open
Description – Control senses lid is open during cycle.
Action (If lid remains closed, but alarm occurs):
1. Faulty LS4 replace.
2. Loose connection at LS4 – repair. (Refer to Figure
5-3.)
3. Incorrect adjustment of latch flag assembly
readjust.
Action (If lid physically opens during cycle):
1. Incorrect latch bar adjustment.
2. Chamber seal leaks replace.
5.1 LIQUID CHEMICAL STERILIZATION CYCLE FAULTS (CONTINUED)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Steris SYSTEM 1E and is the answer not in the manual?

Steris SYSTEM 1E Specifications

General IconGeneral
BrandSteris
ModelSYSTEM 1E
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals