NOTE. You have the option to add user-defined labels for the analog, reference, bus, and digital sources. Digital sources are
only available on MSO models.
Behavior
The instrument acquires data from the parallel bus only when the signal used as the Clock Source changes logic levels.
What do you want to do next?
Learn more about bus setups.
Learn about bus configuration.
Learn about digital setups.
Select the bus clock channel polarity
From the Bus Setup window, select a bus, select the Parallel bus type, and click Clocked.
To use
â–
Connect a probe between the instrument and the clock signal in your system under test.
â–
Click the Clock Polarity selection box and select from Rising, Falling, or Either.
Behavior
â–
Rising causes the instrument to acquire data on the rising edge of the signal.
â–
Falling causes the instrument to acquire data on the falling edge of the signal.
â–
Either causes the instrument to acquire data on the rising and the falling edges of the signal.
What do you want to do next?
Learn more about bus setups.
Learn about bus configuration.
Learn about digital setups.
Bus setups
DPO70000SX, MSO/DPO70000DX, MSO/DPO70000C, DPO7000C, and MSO/DPO5000B Series 153