6.3.13 High Breath Rate Alarm
Property Description
Alarm text
High Breath Rate
Priority Medium
Alarm condition A High Breath Rate alarm will be given when the alarm limit
has been exceeded for 15 seconds.
Possible cause
• Mismatch between the Breath Rate setting and the alarm
setting.
• Increased Breath Rate.
• Too sensitive setting of the inspiratory trigger setting.
Reset criteria
The breath rate goes below the alarm limit.
Ventilator action The Vivo 65 will continue treatment according to the current
settings.
Setting range
• 10 bpm to 70 bpm (Adult mode)
• 10 bpm to 99 bpm (Paediatric mode)
• Off
Setting resolution 1 bpm.
102 Alarms