6.3.15 Apnea Alarm
Property Description
Alarm text
Apnea
Priority High
Alarm condition An Apnea alarm will be given when no patient-triggered breath
is detected for the set period of time.
Possible cause
• Patient stopped breathing.
• Patient decreases spontaneous breathing.
• Circuit disconnection.
• Inspiratory Trigger is set too high.
Reset criteria Inspiratory effort detected by the Vivo 65.
Ventilator action The Vivo 65 will continue treatment according to the current
settings.
Setting range
• 5 to 60 s. (Non MPV mode)
• 15 to 900 s. (MPV mode)
• Off
Setting resolution 5 s below 15 s.
15 s above 15 s.
MPV mode:
15 s below 60 s.
60 s above 60 s.
6.3.16 Disconnection Alarm
CAUTION!
No single alarm can reliably detect all disconnections due to the number of possible
combinations of therapy settings, circuit configurations and patient interfaces. To verify
that patient disconnection can be detected, including if the patient interface becomes
accidentally detached from the patient, it is advised to test the functionality of the Dis-
connection Alarm upfront with the complete set-up as used during treatment, including
items such as filters, circuit, connectors, interface (mask, cannula etc.)
104 Alarms
User manual
Doc. 006150 M-1