Property Description
Alarm text
Vte/MVe Sensor Error
Alarm condition
AVte/MVe sensor error alarm occurs when the sensor for
Vte/MVe measurement fails. The Vte/MVe readings on the
display are not correct.
Contact your service provider. The ventilator can still be used
with single limb circuit.
Priority High
6.4.12 SpO
2
/CO
2
/Remote Start/Stop Failure Alarm (SpO
2
/CO
2
Remote Fail)
Property Description
Alarm text
SpO2/CO2/Remote fail
Alarm condition
An SpO
2
/CO
2
/Remote Start/Stop Failure alarm will be given
when a failure occurs with the patient interface or attached
units.
Priority
Medium
Possible cause
• Failure in the Remote start/stop unit.
• Failure in the SpO
2
sensor.
• Failure in the CO
2
sensor.
• Internal failure in the ventilator.
Ventilator action The ventilator will continue treatment with the same settings.
6.4.13 SpO
2
Disconnected (SpO
2
Sensor Failure/Disconnection Alarm)
Property Description
Alarm text
SPO2 Disconnected
Alarm condition
An SpO
2
Sensor Failure/Disconnection alarm will be given
when an error signal or no signal from the SpO
2
sensor has
been detected for 2 seconds. Check the SpO
2
sensor.
Priority High
Possible cause The SpO
2
electronics cable has been disconnected and subse-
quently no communication (possibly due to disconnection)
for 2 seconds.
Failure in the SpO
2
sensor.
Ventilator action The ventilator will continue treatment with the same settings.
Reset
Confirmation by user or reconnected/changed.
Alarms 117