The new batteries perform best if no complete cycles are made. Therefore it is only an
advantage to charge the battery as soon as an opportunity arises.
The internal and click-on batteries are automatically charged when connecting the Vivo 65
to the mains supply. To ensure that the batteries are fully charged, a maintaining charging
cycle will be performed. The batteries are not charged when connecting the Vivo 65 to an
external DC supply. While charging, the battery level will be animated. The batteries are only
charged if the internal temperatures are between 0 to 45°C (32 to 113°F). High power con-
suming settings in combination with high ambient temperatures may make the battery tem-
perature rise above 45°C (113°F).
Charging Times
BATTERY CHARGER TIME*
Internal battery
Vivo 65 3 h
Click-on
battery
Vivo 65 5.5 h
Click-on
battery
Click-on battery charger 3 h
* Times are based on charging empty batteries.
5.7.3 Battery Icons
When running on battery, the battery status is indicated by the following symbols:
SYMBOLS BATTERY STATUS
Full
Medium
Empty/Low
Disconnected or malfunctioning
Malfunctioning
5.7.4 Internal Battery
The internal battery is intended as a backup power source if the primary AC power (Mains)
source fails. It can also be used as a temporary power source. For example during transporta-
tion between one stationary power source to another.
72 How to Use the Vivo 65
User manual
Doc. 006150 M-1

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Breas Vivo 55 and is the answer not in the manual?

Breas Vivo 55 Specifications

General IconGeneral
BrandBreas
ModelVivo 55
CategoryMedical Equipment
LanguageEnglish

Related product manuals