Volume Control Modes
The worst case Vivo 65 System is the dual limb patient circuit with or without HCH
humidifier, bacterial filter, FiO
2
sensor and CO
2
sensor.
8.2 Data Parameters
NOTE
All stated tolerances includes measurement uncertainty. The accuracies have been tested
with all allowed configurations. Stated tolerances only discloses the maximum tolerance.
If a parameter’s tolerance is described with both absolute and relative measures, the
greater one applies if nothing else is stated. Percent measurements are relative to the set
value.
8.2.1 Settings
Please refer to the information in 5.4 Functions and Parameters in the Vivo 65, page 56.
8.2.2 Monitored Values
MONITORED
VALUE
RANGE ACCURACY
P
peak
4 to 70 cmH
2
O.
±0.5 cmH
2
O or ±10%, which-
ever is greatest
PEEP 0 to 30 cmH
2
O.
±0.5 cmH
2
O or ±10%, which-
ever is greatest
P
mean
0 to 70 cmH
2
O.
±0.5 cmH
2
O or ±10%, which-
ever is greatest
Leakage
0 to 100 l/min (BTPS*).
±10%
MV
i
0 to 99.9 l (BTPS*). ±10% or (±10 ml × bpm),
whichever is greatest
MV
e
0 to 99.9 l (BTPS*). ±10% or (±10 ml × bpm),
whichever is greatest
Vt
i
0 to 9999 ml (BTPS*). ±10 ml or 10%, whichever is
greatest
Vt
e
0 to 9999 ml (BTPS*). ±10 ml or 10%, whichever is
greatest
FiO
2
0 to 100%.
±2%**
% in TgV 0 to 100%.
±1%
Total Rate 0 to 99 bpm. 1 bpm
Spont Rate 0 to 99 bpm. 1 bpm
134 Technical Specifications
User manual
Doc. 006150 M-1