Property Description
Priority
Medium
Ventilator action The ventilator will continue treatment with the same settings.
6.4.5 Lost Mains Alarm
Property Description
Alarm text
Lost Mains Power
Alarm condition A Mains Power Lost alarm will be given when the ventilator
switched from AC power (Mains) to another power source due
to AC Power (Mains) is lost.
Priority Medium
Ventilator action The ventilator will continue treatment with the same settings.
An information message will be shown on the screen.
Reset
Confirmation by user or AC power (Mains) reconnected.
6.4.6 Crit. Low Last Power Source Alarm
Property Description
Alarm text
Crit. Low Last Power Source
Alarm condition A Crit. Low Last Power Source alarm will be given when the
last battery source (internal battery or click-in battery) has
5 minutes of operating time left with current settings.
Priority High
Ventilator action The ventilator will continue treatment with the same settings.
Reset
Connection of “higher” power source.
114 Alarms
User manual
Doc. 006150 M-1