Cooling Air Inlet Filter
The cooling air inlet filter is located in the filter cassette at the side of the ventilator. Replace
the filter at least once a year. Wash the filter at least once a week.
1 Wash the filter using warm water and a
mild soap.
2 Rinse thoroughly.
3 Dry the filter by squeezing it out in a towel. Do not wring the filter.
4 Make sure the filter is completely dry before inserting.
7.1.2 Insert
Always use a new dual insert when the ventilator is to be used by a new patient.
Check the insert regularly for damage. In case of damage, replace the insert.
CAUTION!
Appropriate personnel should determine the duration of use for the insert based on
accepted infection control procedures.
7.1.3 Patient Circuit
The patient circuit should be cleaned and replaced in accordance with the manufacturer’s
instructions and care provider’s instructions, where applicable.
Always use a new patient circuit when used by a new patient.
Check the patient circuit regularly for damage. In case of damage, replace the circuit.