6.3.20 Low FiO
2
Alarm
Property Description
Alarm text
Low FiO2
Priority High
Alarm condition A Low FiO
2
alarm will be given when the measured FiO
2
is
below the alarm limit for 30 seconds.
Possible cause
• Decreased oxygen inlet.
• Disconnection of oxygen inlet.
• Increased minute ventilation.
• High leakage.
Setting range
• 21% to 100%
• Off
Setting resolution 1%
Ventilator action The ventilator will continue treatment with the same settings.
6.3.21 High SpO
2
Alarm
Property Description
Alarm text
High SpO2
Priority
Medium
Alarm condition A High SpO
2
alarm will be given when the measured SpO
2
exceeds the alarm limit for 30 seconds.
Possible cause Too high flow of bleed-in oxygen.
Reset criteria
The SpO
2
value goes back below the alarm limit.
Ventilator action The Vivo 65 will continue treatment according to the current
settings.
Setting range
• 90 % to 100 %
• Off
Setting resolution 1 %
This alarm requires a connected SpO
2
sensor.
108 Alarms
User manual
Doc. 006150 M-1