EasyManuals Logo

Breas Vivo 55 User Manual

Breas Vivo 55
165 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #53 background imageLoading...
Page #53 background image
In the Alarm section, the alarm parameters can be altered. (See 6 Alarms, page 88 for more
information .)
The Alarm/Event History screen displays all events that have been logged, as well as alarms
that have occurred. Alarms are coloured according to priority, and are maintained when the
Vivo 65 is powered down.
The manufacturer-configured state of physiological alarms may be retrieved by selecting
“Reset to basic settings” on the Mode screen in Standby mode.
5.2.8 The Monitor Section
The monitoring section provides display of treatment data.
The monitoring screen contains a bargraph field that displays current Pressure, PEEP and
Ppeak, pressure and volume alarm limits, and inspired/expired tidal volume.
The monitoring field displays all available values for the current treatment mode. (In most
other screens, except for the curve and trend view, a small monitoring field is displayed with
8 values.) See 5.3 Monitored Values in the Vivo 65, page 53 for a description on the monitored
values.
5.2.9 The Others Section
WARNING!
For the ventilator to deliver treatment according to settings, it is important that the selec-
tion of the patient circuit type is correctly set.
NOTE
The currently selected profile cannot be turned off.
If changing between Adult and Paediatric modes, the active profile is reset.
How to Use the Vivo 65 51
Doc. 006150 M-1 User manual

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Breas Vivo 55 and is the answer not in the manual?

Breas Vivo 55 Specifications

General IconGeneral
BrandBreas
ModelVivo 55
CategoryMedical Equipment
LanguageEnglish

Related product manuals