5.3.11 Total Rate
The Total Rate displays the actual total breath rate independent of whether the breaths are
patient- or ventilator-triggered breaths.
5.3.12 Spont Rate
The Spont Rate displays the actual spontaneous breath rate.
5.3.13 % Spont
% Spont displays the percentage of spontaneous breaths calculated since the ventilator was
last started (not calculated until 100 breaths are registered).
5.3.14 SpO
2
(Oxygen Saturation)
The SpO
2
displays the patient’s oxygen saturation, if measured with the SpO
2
module
accessory.
• When using the ventilator with the SpO
2
sensor accessory, the ventilator displays func-
tional oxygen saturation measured by the sensor.
• The following information concerns the light emitted by the SpO
2
Sensor:
1. Peak Wavelength (red): 660 nm
2. Peak Wavelength (infrared): 905 nm
3. Maximum Optical Output Power: 15 mW
4. For more information regarding the oxygen probe’s range of peak wavelengths, max opti-
cal power and usage, please refer to the respective probe manual.
• Environmental factors may influence the function or accuracy of the pulse oximeter,
such as ambient light, physical movement, diagnostic testing, low perfusion, electromag-
netic interference, dysfunctional haemoglobin, presence of certain dyes and inappropri-
ate positioning of the pulse oximeter probe.
• A functional test cannot be used to assess the Accuracy of a Pulse Oximeter Probe or a
Pulse Oximeter.
5.3.15 Pulse Rate
Pulse Rate displays the patient’s pulse rate as measured with the SpO
2
module.
5.3.16 I:E
I:E displays the ratio between the length of the inspiration and the length of the expiration.
5.3.17 PScalc
PScalc displays the pressure above the PEEP pressure applied to the patient in PCV/PSV
mode.
5.3.18 Insp. Time
Insp. Time displays the duration of the inspiratory cycle, measured from the start of inspira-
tion to the start of expiration.
5.3.19 Rise Time
Rise Time displays the duration of the pressure or volume increase, measured from the start
of inspiration until the set pressure or volume has been reached.
How to Use the Vivo 65 55
Doc. 006150 M-1 User manual