6.4.17 CO
2
Sensor Error Alarm
Property Description
Alarm text
CO2 Sensor Error
Alarm condition A CO
2
Sensor Error alarm will be given when an error in the
CO
2
sensor has occurred.
Replace the CO
2
sensor. CO
2
monitoring cannot be performed
in this condition.
Priority High
Ventilator action The ventilator will continue treatment with the same settings.
Reset
OK message from sensor or sensor disconnected by user.
6.4.18 FiO
2
Disconnected (FiO
2
Sensor Failure/Disconnection Alarm)
Property Description
Alarm text
FiO2 Disconnected
Alarm condition
An FiO
2
Sensor Failure/Disconnection alarm will be given
when no signal from the FiO
2
sensor has been detected for
2 seconds. Check the FiO
2
sensor.
Priority High
Possible cause
• FiO
2
Sensor disconnected.
• Communication with the FiO
2
sensor failed.
Ventilator action The ventilator will continue treatment with the same settings.
Reset
Confirmation by user or reconnected/changed.
Alarms 119
Doc. 006150 M-1 User manual