6.3.18 Obstruction Alarm
Property Description
Alarm text
Obstruction
Priority High
Alarm condition An Obstruction alarm will be given if the inspiratory breathing
tube becomes blocked and remains blocked for 2 consecutive
breaths.
Note: An obstruction of the expiratory breathing tube can be
detected with the Low Vt
e
alarm.
Ventilator action With each breath cycle, upon detection of an obstruction the
ventilator will reduce the airway pressure to the set PEEP.
Treatment will resume with the start of the next breath cycle.
Reset Criteria
When the monitored compliance and resistance become nor-
mal after a breath.
Setting Range
High
Low
Off
6.3.19 High FiO
2
Alarm
Property Description
Alarm text
High FiO2
Priority
Medium
Alarm condition A High FiO
2
alarm will be given when the measured FiO
2
exceeds the alarm limit for 30 seconds.
Possible cause
Increased oxygen inflow.
Decreased minute ventilation.
Reset criteria
FiO
2
goes below the alarm limit
Setting range
21% to 100%
Off
Setting resolution 1%
Ventilator action The ventilator will continue treatment with the same settings.
Alarms 107
Doc. 006150 M-1 User manual

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Breas Vivo 55 and is the answer not in the manual?

Breas Vivo 55 Specifications

General IconGeneral
BrandBreas
ModelVivo 55
CategoryMedical Equipment
LanguageEnglish

Related product manuals