6.3.14 Low Breath Rate Alarm
Property Description
Alarm text
Low Breath Rate
Priority High
Alarm condition A Low Breath Rate alarm will be given when the delivered total
breath rate is below the alarm limit for 15 seconds.
Possible cause
Mismatch between the Breath Rate setting and the alarm
setting.
The patient cannot trigger breaths because the inspiratory
trigger setting is too high.
Decrease in the patient’s spontaneous breathing.
Circuit disconnection.
Reset criteria
The breath rate goes above the alarm limit.
Ventilator action The Vivo 65 will continue treatment according to the current
settings.
Setting range
Adult: 1 bpm to 30 bpm.
Paediatric: 4 bpm to 50 bpm.
Off
Setting resolution 1 bpm.
Alarms 103