6.4.24 Alarm for Beeper Failure
Property Description
Alarm text
Alarm Beeper Fail
Alarm condition Failure of the beeper control by the treatment processor.
Priority High
Ventilator action The ventilator will continue treatment with the same settings.
6.4.25 Int. Function Failure (Internal Function Failure Alarm)
Property Description
Alarm text
Int. function failure:
Alarm condition An Internal Function Failure alarm will be given when the ven-
tilator has an internal function failure, followed by an error
code for the specific failure. All Internal Function Failure alarm
error codes are defined and explained in the ventilator Service
Manual.
Priority High
Ventilator action The ventilator will stop the treatment and shut down.
Ventilator reset Restart the ventilator.
6.5 Alarm Test
6.5.1 Instructions
NOTE
• In order to test an alarm it must be activated by your clinician.
• In some cases a setting must be altered.
To perform the alarm test, follow the instructions below:
Low Pressure and Disconnection Alarms
1 Start treatment and disconnect the patient circuit.
2 Wait 15 seconds.
3 The Low Pressure Alarm and/or the Disconnection Alarm will be given.
4 Stop treatment. Test completed.
Low Vti or Low Vte Alarms
1 Start treatment and block the patient circuit completely to simulate an occlusion.
2 Wait 15 seconds.
122 Alarms
User manual