6.3.3 High PEEP Alarm
Property Description
Alarm text
High PEEP
Priority Medium
Alarm condition A High PEEP alarm will be given when the measured PEEP is
30% above the set value for more than 15 seconds
Possible cause
Blocked leakage port.
Too short expiratory time.
Changes in airway resistance and or compliance.
Malfunction of the exhalation valve.
Blocked exhalation valve.
Reset criteria
PEEP has gone below the alarm limit (lower than 30% above
the set value).
Ventilator action The Vivo 65 will continue treatment according to the current
settings.
Setting range
On
Off
6.3.4 Low PEEP Alarm
Property Description
Alarm text
Low PEEP
Priority
Medium
Alarm condition A Low PEEP alarm will be given when the measured PEEP is
30% below the set value for more than 60 seconds
Possible cause
Excessive leakage.
Malfunction of the exhalation valve.
Reset criteria
PEEP has gone above the alarm limit (higher than 30% below
the set value).
Ventilator action The Vivo 65 will continue treatment according to the current
settings.
Setting range
On
Off
Alarms 93
Doc. 006150 M-1 User manual

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Breas Vivo 55 and is the answer not in the manual?

Breas Vivo 55 Specifications

General IconGeneral
BrandBreas
ModelVivo 55
CategoryMedical Equipment
LanguageEnglish

Related product manuals