MONITORED
VALUE
RANGE ACCURACY
% Spont 0 to 100%.
Not applicable
SpO
2
70 to 100%.
±3 digits. Data update period 1
s.
4-beat average signal processing.
Pulse Rate 18 to 300 bpm. ±3 digits. Data update period 1
s.
4-beat average signal processing.
I:E 1:99 to 10:1.
±0.1 unit
Insp. Time 0.3 to 5 s.
±0.1 s
Rise Time 0.1 to 5 s.
±10% or ±0.1 s, whichever is
greatest
EtCO
2
0 to 25%.
0 to 15%: ±(0.2 vol% + 2% of
reading), 15 to 25%: unspecified
InspCO
2
0 to 25%.
0 to 15%: ±(0.2 vol% + 2% of
reading), 15 to 25%: unspecified
* BTPS (Body Temperature and Pressure Saturated)
** Effects of humidity on FiO
2
accuracy: -0.03% per % RH relative to calibration point
8.2.3 Alarms
Please refer to the information in 6 Alarms, page 88.
8.2.4 Technical Specifications and Conditions
8.2.4.1 Power Supply
POWER SUPPLY
SPECIFICATION
AC Power (Mains) supply
100 to 240 V AC, tolerance: +10%/-20%, 50 to 60
Hz, max 300 VA.
External battery 24 V DC, tolerance: 24 V ± 6 V.Max 7 A, 140 W.
Click-on battery* Capacity 5.2 Ah. LiIon. Operational time 8 hours, life-
time 3 years.
Internal battery Capacity 2.6 Ah. LiIon. Operational time 4 hours, life-
time 3 years.
*For transportation by air, please observe that the click-on battery capacity is 192 Wh which
exceeds stated limits. Always consult your airline company for transportation restrictions.
Technical Specifications 135
Doc. 006150 M-1 User manual