Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n 107
Getting started
6.6 Checking the backup manual switch
When starting up for the first time, check that the backup manual switch is
functioning.
Prerequisite: The device is in standby mode.
1. Open the flap (1).
2. Activate the backup manual switch (2).
An optical alarm signal (3) with high priority is issued. A dialog (4) is also
displayed.
If the alarm and dialog are not displayed, contact specialized service
personnel.
3. Deactivate the backup manual switch again.
5517955180
2
1
4
3
Exit backup manual mode