Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n 185
Operation
2. Touch the System > Logbook tab (1).
3. Touch the appropriate button (2).
8.13.3 Resetting user-specific settings
Changes made in the System setup dialog during operation can be reset to the
default settings.
1. Open the System setup dialog.
2. Open the corresponding page.
3. Touch the System defaults button and confirm.
8.14 Ending the therapy
8.14.1 Switching to the standby mode
1. Touch the Standby... button in the main menu bar.
Or
Press the key below the screen.
2. Confirm with the rotary knob.
WARNING
Soda lime drying out
Moisture losses occur if fresh gas is continuously passed through the soda lime. If
the moisture falls below the minimum level, the following adverse reactions occur
regardless of the type of soda lime and inhalational anesthetic agent used:
– Reduced CO
2 absorption and consequently an increase in inspiratory CO2
values
– Increased generation of heat in the CO
2 absorber and consequently increased
breathing gas temperature
– Formation of carbon monoxide
– Absorption and/or degradation of the inhalational anesthetic agent
► With mechanically controlled gas mixers:
Do not leave the flow control valves open for an unnecessarily long period of
time.
35981
1
2
1