326 Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
Principles of operation
Changing the body weight during mechanical ventilation has no effect on the
current ventilation settings.
16.7.3 Influence of patient age
During operation, the set age influences the following:
– Calculation of MAC value
16.8 Support of Infinity ID accessories
The following accessories can be used if the device has the appropriate options:
– Infinity ID breathing circuit
– Infinity ID WaterLock 2 water trap
– Infinity ID flow sensors
– Infinity ID CLIC absorber
The Infinity ID functionality can be configured, see page 221.
16.8.1 Infinity ID functionality
16.8.1.1 Replacement interval monitoring
Automatic monitoring of the period of use is available for Infinity ID products. An
exceeded period of use is signaled during the system test.
The replacement interval for the connected Infinity ID accessories can be adjusted.
This interval must be specified in accordance with the applicable infection
prevention regulations or the requirements stated in the instructions for use for the
corresponding accessory.
16.8.1.2 Anti-interchange security
When Infinity ID breathing hoses and Infinity ID breathing bags are used, the
incorrect connection of breathing hoses and the breathing bag is detected and
reported. Hoses that are incorrectly connected with the breathing system trigger an
automatic alarm.
16.9 Schematic illustration of the acoustic signals
16.9.1 Alarm signal for various alarm priorities
Alarm priority Standard (according to IEC 60601-1-8) Repeated
High
Depending on the overall alarm situation,
this acoustic alarm signal may also be
played as only a 5-tone sequence due to
the timing of the individual alarms.
Yes
Medium Yes
Low No