Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n 75
Assembly and preparation
4.4 Intrahospital transport
Transport includes any movement of the device to other rooms or functional areas.
Alignment or positioning of the device within the operating room is not counted as
transport.
4.4.1 Increasing the tipping stability during transport
1. Carefully fold the arms with any mounted accessories against the device, (e.g.,
patient monitor, data management system, syringe pumps). Insert the additional
pull-out work surface. Fold down the additional folding table extension.
2. Use a strap to prevent the arms from swinging out in an uncontrolled manner.
3. Remove all loose objects from the attached arms and the shelves.
4. Remove all objects from the standard rails.
5. Remove the vaporizers.
6. Remove all objects from the writing tray and the work surface.
7. Slide the additional pull-out writing tray (if present) completely into the device.
8. Position the breathing bag arm (if present) close to the device.
9. Push in and lock the drawers.
4.4.2 Parking the medical device
Always apply all brakes when parking the device, especially on inclined surfaces.
4.4.3 Visual inspection after transport
WARNING
Risk of device malfunction
The device, especially hoses and cables, can become damaged during transport.
As a result, the device function may be impaired. The patient may be put at risk.
► Check the device for damage after transport, particularly the hoses and cables.
1. Check the device for damage, particularly the hoses and cables.
2. Do not operate a damaged device. The damage must be repaired by service
personnel before the device is used.
3. Before connecting the device to the power supply, check the electrical
connections and auxiliary power sockets for residual moisture. Wipe off the
moisture. Ensure sufficient drying time.
4. Before operating the device, make sure it is on a level surface without
inclination.