108 Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
Getting started
6.7 Emergency start-up
6.7.1 Mechanically controlled gas mixer with flow tubes
1. Adjust the APL valve.
2. Open the flow control valve (1) and set the desired O
2 flow. If necessary, press
the O
2+ key (3) to quickly fill the breathing bag.
3. Monitor the flow on the flow tubes (2).
4. Set the anesthetic gas concentration on the vaporizer.
5. Ventilate the patient manually.
6. Turn on the device.
7. As soon as the Standby page is displayed, start the therapy, see page 143.
6.7.2 Mechanically controlled gas mixer with electronic flow
measurement
1. Adjust the APL valve.
53614
1
2
3
35054