Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n 267
Service
13 Service
13.1 Safety instructions
Intervals and implementation
Wear and material fatigue of the components may lead to device failure and
malfunctions. If services activities are not performed regularly or properly,
malfunctions may occur, which can result in personal injury and property damage.
► Perform service at the specified intervals.
► Service activities must be performed by user groups that are assigned to the
specific activity.
► Only perform service activities when there is no patient connected to the device.
► Before performing maintenance, disconnect all electrical connections from the
power supply and all gas connections from the gas supply.
► Perform a system test after service activity.
Housing
Under the housing, there are live electrical components, which may cause an
electric shock.
► The housing may only be opened by user groups that are assigned to the
specific activity.
Risk of infection
If the product has not been reprocessed properly, it may be contaminated with
pathogens. As a result, persons could be put at risk.
Perform the following measure before servicing, before returning the device, and
before disposal:
► Perform the reprocessing in accordance with the reprocessing instructions
supplied with the product.
13.2 Definition of service terminology
Concept Definition
Service All measures (inspection, maintenance, repair) intended to
maintain or restore the functional integrity of a product.
These measures can be supported using the connected ser-
vice functionality (data exchange between the product and
DrägerService via a network). For further information see:
"LAN", page 303.
Inspection All measures intended to assess the current state of a product
Maintenance Regular specified measures intended to maintain the functional
integrity of a product
Repair Measures intended to restore the functional integrity of a prod-
uct after a failure