268 Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
Service
13.3 Inspection
The maximum scope for all main devices and options is listed below. Observe the
equipment that is applicable for the present device.
Safety checks
Safety checks are not a substitute for maintenance, which includes the preventive
replacement of wearing parts as specified by the manufacturer.
Performing the safety checks:
1. Check that the associated instructions for use are available in the correct
version.
2. Perform a functional test of the following functions in accordance with the
instructions for use:
–Emergency O
2 delivery
3. Check that the product is in good condition:
– All labels are complete and legible
– No visible damage to the following components:
– Trolley and castors
– Housing parts
– Brakes
– Vaporizer mount
– Water trap holder
– O-rings of the water trap holder
– Sample line
– Screen
– Gas inlets
– Status display
– Breathing system
– Piston drive and piston diaphragm
– AGS and AGS valves
– Hoses and cables
– Strain relief for compressed gas hoses and cables
– Support arms
– Pressure reducers and their sensor cables
– Fuses that are accessible from the outside are in compliance with the
specified values.
– Check that the country-specific labeling of the gas type matches the screen
display.
– Check the end stops of the support arms for patient monitors.
4. Check the electrical safety in accordance with the IEC 62353 standard.
Measure Interval User group
Inspection and safety check Every 12 months Service personnel