182 Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
Operation
8.11 Changing the patient data
Patient data can be changed during operation.
1. Open the Patient dialog.
2. Modify the patient data.
Changes influence the therapy suggestions, among others, which is indicated by
the position of the arrow on the therapy controls.
The current therapy settings remain unaffected. Observe the following information:
"Influence of patient category, weight, and age on device behavior", page 325.
When the patient category is changed, the age, weight, and height are
automatically adapted so that they remain within the defined limits, see page 210.
8.12 Exporting data
8.12.1 General information
This device enables the export of screen contents, trends, and data on a USB mass
storage device.
Prerequisites:
– The USB port is activated in the system setup.
– A USB mass storage device is connected to the USB port.
During a saving process, the button turns dark green.
The data are stored in the "Draeger\ExportData" directory.
8.12.2 Exporting the screen contents
The screen contents can be exported to a USB mass storage device as a
screenshot.
● Touch the Export screenshot button in the main menu bar.
The screenshot will be saved as a ".png" file.
35978