Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n 239
Troubleshooting
2. On the underside, remove the water trap (2) and the RFID antenna, if present.
To do this, gently press the side surfaces (3) of the RFID antenna together. Pull
the RFID antenna off downwards and leave it on the device.
Do not pull on the antenna cable. Do not kink the antenna cable.
3. Remove the 2 screws and the fixing plate (4).
4. Withdraw the patient-gas measurement module from the compartment.
5. Fit a new patient-gas measurement module. Proceed in reverse order to do this.
6. Insert a new water trap.
7. Turn on the device.
8. Perform a software download for the new PGM if required.
9. Perform a system test.
12.8 Flow measurement failure
If the flow measurement fails, mechanical ventilation can still continue. There may
be limitations with regard to displayed measured values, measurement accuracies,
and when triggering mandatory breaths. The flow sensors can be replaced as soon
as the device is in standby mode.
1. Replace the flow sensors.
2. Perform the leakage test, see page 112.
40837