188 Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
Operation
WARNING
Risk of fire
In combination with oxygen or nitrous oxide, ignition sources such as
electrosurgical devices and laser surgical devices can cause fires.
► If ignition sources are present, do not open the flow control valve on the O
2
flowmeter. Leave the flow control valve completely closed.
3. Open the flow control valve (2) to begin O
2 insufflation.
Ending O
2 insufflation:
● Close the flow control valve (2).
8.15 Change of patient
8.15.1 Cleaning and disinfecting the workstation
● Clean and disinfect the workstation in accordance with the infection prevention
policy of the health-care facility. Observe the reprocessing instructions supplied
with the device.
8.15.2 Checking or replacing consumables
Prerequisite: The device is in standby mode.
Integrated patient-gas measurement module
WARNING
Risk of infection
Used sample lines may be infectious due to the breathing gases that passed
through them.
► If the sample line is connected directly to the Y-piece without a filter, replace the
sample line after every patient.
► Replace the sample line regularly. Follow the infection prevention policies and
reprocessing regulations of the health-care facility.
1. If no filter was used, replace the sample line and dispose of the used sample
line.
WARNING
Risk of infection
The water trap may contain infectious fluid.
► In order to prevent infectious fluid from spurting out, first remove the sample line
from the water trap and perform surface disinfection.
► Empty or replace the water trap according to its instructions for use.
► Follow the infection prevention policies and reprocessing regulations of the
health-care facility.
2. Check the water trap of the patient-gas measurement module (PGM) for
leakage. If necessary, empty or replace the water trap.
Vaporizer filling level
● Check the vaporizer filling level in the sight glass. Fill the vaporizer if required.