270 Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
Service
13.4 Maintenance
The maximum scope for all main devices and options is listed below. Observe the
equipment that is applicable for the present device.
It is recommended that only original parts from Dräger are used and that the parts
are replaced by DrägerService.
A service contract with DrägerService is recommended.
13.5 Repair
Repairs may be performed only by specialized service personnel.
It is recommended that only original parts from Dräger are used and that the parts
are replaced by DrägerService.
A service contract with DrägerService is recommended.
Component Interval Measure User group
CO
2 absorber/soda lime
with disposable dust fil-
ter
When colored violet or
according to the config-
ured Infinity ID replace-
ment interval
Replace, see page 93 Users
Water trap As required, when
soiled, or according to
the configured Infinity ID
replacement interval
Replace, see page 95 Users
Flow sensors As required, if calibra-
tion is no longer possi-
ble, or according to the
configured Infinity ID
replacement interval
Replace, see page 240 Users
Piston diaphragm Annually
Replace
1)
1) The replacement is described in the separate reprocessing instructions for the anesthesia machine. The reprocessing
instructions are included in the delivery of the anesthesia machine.
Users
O
2 sensor cell Every 2 years Replace, see page 69 Users
AGS As required, when
soiled
Replace Specialized service per-
sonnel
AGS particle filter If required, if filter is
soiled, or flow is no lon-
ger achieved
Replace Users
O-rings on the water
trap holder
Every 2 years Replace Service personnel
Pressure reducers Every 6 years Replace Specialized service per-
sonnel
Filter cloth
– Above the gas mixer
– Power supply unit
Every 2 years Replace Service personnel
CLIC adapter Every 4 years Replace Users
Lead battery (2 pieces) Every 2 years Replace Service personnel