300 Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
Technical data
15.24.3 Recommended separation distances from wireless
communication devices
To ensure that the technical data of this device can be met, there must be a
separation distance of at least 1.0 m (3.3 ft) between this device and wireless high-
frequency communication equipment.
15.25 Device combinations
This device can be operated in combination with other Dräger devices or with
devices from other manufacturers. Follow the accompanying documents of the
individual devices.
If a device combination is not approved by Dräger, the safety and the functional
integrity of the individual devices can be compromised. The operating organization
must ensure that the device combination complies with the applicable editions of
the relevant standards for medical devices.
Device combinations approved by Dräger meet the requirements of the following
standards:
IEC 60601-1, 3rd edition (general requirements for safety, device combinations,
software-controlled functions)
IEC 60601-1-2: 3rd edition (electromagnetic compatibility) or 4th edition
(electromagnetic interference)
IEC 60601-1-8 (alarm systems)
Or:
IEC 60601-1, 2n edition (general requirements for safety)
Immunity against Test level and required electromagnetic environ-
ment
Electrostatic discharge (ESD)
(IEC 61000-4-2)
Contact discharge: ±8 kV
Air discharge: ±15 kV
Fast transient electrical disturbances (bursts)
(IEC 61000-4-4)
Power cable: ±2 kV
Longer signal input lines/output lines: ±1 kV
Impulse voltages (surges)
(IEC 61000-4-5)
Voltage, external conductor – external conductor:
±1 kV
Voltage, external conductor – protective ground con-
ductor: ±2 kV
Magnetic fields at mains frequency
(IEC 61000-4-8)
50 Hz: 30 A/m
Voltage dips and short interruptions in the
supply voltage (IEC 61000-4-11)
Voltage dips of 30 % to 100 %, 8.3 ms to 5 s,
different phase angles
Radiated high-frequency disturbances
(IEC 61000-4-3)
80 MHz to 2.7 GHz: 3 V/m
Conducted high-frequency disturbances
(IEC 61000-4-6)
150 kHz to 80 MHz: 3 V, ISM bands: 6 V
Electromagnetic fields in the vicinity of wireless
communication devices
Various frequencies from 385 MHz to 5785 MHz:
9 V/m to 28 V/m

Table of Contents

Other manuals for Dräger Atlan A100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Atlan A100 and is the answer not in the manual?

Dräger Atlan A100 Specifications

General IconGeneral
BrandDräger
ModelAtlan A100
CategoryMedical Equipment
LanguageEnglish

Related product manuals