Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n 143
Operation
8.1.6 Exporting data
Unauthorized access to the device can impair the device function. When using USB
mass storage devices, a denial of service may occur.
► Only devices that are included in the list of accessories or which correspond to
the "mass storage medium" USB device class may be connected. For example,
no provision has been made for connecting devices whose battery is to be
charged.
► Dräger recommends the use of storage media with hardware encryption.
USB mass storage devices must be checked for malware and approved before
being connected to the device. Checking and approval must be performed by
the IT administrator or equipment officer of the health-care facility in accordance
with the specifications of the health-care facility.
The following USB devices must not be connected:
– Active USB devices (own power supply)
– Devices that use wireless technology
– Cellular phones
8.2 Starting the therapy
The therapy can be started either from a quick start into the Manual / Spontaneous
mode or from a normal start with customized settings.
Prerequisite: The device is in Standby mode.
Quick start
Mechanically controlled gas mixers:
● Touch the Quick start Man/Spon button (1) and open the flow control valves.
35280