7.9.15 Diaphragm brakeage
An automatic monitoring of the pump diaphragm can be switched on or of
f under
‘Diaphragm brakeage’ .
Requirements:
n This function can only be used when a pump head of 30 l/h, 50 l/h or 120 l/h size is
installed on the pump with the respective diaphragm protection sensor
.
n Select the corresponding sensor for diaphragm breakage
Ä
Chapter
12.2.2.3 ‘Replacement part module: Pump head 30 l/h and 50 l/h’ on page 182, pos. 9
or
Ä
Chapter 12.2.2.4 ‘Replacement part module: Pump head 120 l/h’ on page 183,
pos. 11 and install.
Electric connection - Sensor for diaphragm breakage
The electric connection is done on the main board
Ä
Chapter 6.3.4.1 ‘Terminal assignment’ on page 52
The connections are as follows:
14 = GND
15 = Diaphragm
16 = 5 V
olt
17 = Anode
Switch on monitoring for diaphragm breakage:
Fig. 72: Switch on monitoring for diaphragm breakage
Factory setting: Monitoring for diaphragm breakage switched off
1.
[Call up Main menu] :
Ä
Chapter 7.6 ‘Main menu’ on page 66
2.
[Select Configuration] :
Ä
Chapter 7.9 ‘Overview - Configuration’ on page 83
3. Use the scrollbar to scroll to the [Diaphragm brakeage] menu option.
4. [Select Diaphragm brakeage] .
ð
Screen switches to the query ‘with diaphragm breakage.’.
5. If required, ‘select with diaphragm breakage.’ .
6.
Pressing the button saves the setting.
ð
The screen switches to the overview: [Configuration].
7.
By pressing the
button, you can return to the [Main menu].
8.
By pressing the
button, you can return to the [operating display].
Control / software
114Rev. 2-05.2018