EasyManuals Logo
Home>GE>Test Equipment>ADTS542F

GE ADTS542F User Manual

GE ADTS542F
153 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #127 background imageLoading...
Page #127 background image
ROUTINE MAINTENANCE
K0553 Revision D 126 [EN] English
Routine Maintenance
Absolute cleanliness of the work area, tools and equipment are essential.
Replacing the output connector o-ring
After inspection as detailed in maintenance task B, carry out the following if the o-ring is worn or
damaged:
Figure 5-1: Replacing the o-ring
1) Carefully remove the o-ring from the small groove at the top of the connector. Fit a new
o-ring in the small groove at the top of the connector.
2) Make sure the o-ring is tight in the groove and not damaged after fitting.
NOTE: Damage to this o-ring causes leaks.
Replacing a fuse
1) Disconnect the electrical power supply.
2) Unscrew the fuse holder cap and remove the fuse.
3) Only fit a new fuse of the correct type and rating into the fuse holder cap.
4) Secure the fuse holder in the front panel.

Table of Contents

Other manuals for GE ADTS542F

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE ADTS542F and is the answer not in the manual?

GE ADTS542F Specifications

General IconGeneral
BrandGE
ModelADTS542F
CategoryTest Equipment
LanguageEnglish

Related product manuals